As an electrical engineer, I distinguish between two kinds of regulation. Which brings me to my fourth observation from the October “AI Transformation of Healthcare” conference: regulation.

During the closing panel, I must have heard five times that regulation is not really a problem. Quite the opposite, it is a benefit; in other fields we would suffer without it. Hard to disagree. It is just that every time, I missed the other side of the coin. If it is not such a problem, how exactly does it not weigh on development, in time or in cost? That answer never came.

Technical regulation rests on physics

Technical regulation starts from how electric current acts on the human body and how earthing, insulation and circuit protection keep us safe. A good example is the safety standard for medical electrical equipment, EN 60601-1. Patient leakage current from an ordinary applied part is limited to 100 microamps, or 500 under a single fault. For a part that may come into contact with the heart, the limit is ten times lower, because there even microamps can trigger fibrillation. Protective earth may have at most 0.1 ohm and is tested with a current of 25 amps. A surface that is not meant to heat the patient may reach no more than 41 °C.

Each of those numbers has a clear reason behind it and is checked with an instrument. Pass or fail. No interpretation, no haggling over wording. And above all, every such requirement directly lowers the chance that someone gets hurt. When the insulation survives the test and the leakage current stays under the limit, the patient is safer. Not on paper, but physically.

Administrative regulation rests on processes

On the other side is what the MDR asks of medical device software. Risk class, quality management system, risk management, software life cycle, clinical evaluation, technical documentation, notified body, post-market surveillance. Under Rule 11, software that provides information for diagnostic or therapeutic decisions usually lands in class IIa at least, and with that, almost always in front of a notified body. For software that is supposed to keep improving, this hurts twice: a substantial change to an algorithm that can affect diagnosis or treatment is typically assessed again. Not every update, but precisely the ones that matter.

None of this has a threshold like 100 microamps. Conformity is shown through documentation and audits. It reduces risk too, only indirectly: because the manufacturer has a process in place, keeps records, checks and takes responsibility.

Technical regulation is proven by measurement, administrative regulation by documentation and audit. Both have their place; they simply rely on a different kind of evidence.

What the paperwork costs

Now let me be quite frank. Every hour and every euro that goes “into paper” is missing from the technical solution. From a better test, a better model, validation with clinicians. And that applies to software too.

The numbers do exist; they just were not mentioned on the panel. In a MedTech Europe survey from 2024, manufacturers reported an average of 19.5 months for the quality system assessment and 21.8 months for the technical documentation assessment. Of the cost of getting certified, 90% goes on their own staff’s work; notified body fees make up only 7%. In a study for the European Commission with data as of October 2025, the total cost of a first product certificate averaged roughly €195,000, with a median of €100,000 and a range from €15,000 to €1.3 million. These are not figures for a single app but averages across the industry. Still, the order of magnitude is a year and a half to two years, and hundreds of thousands of euros.

In the same survey, almost half of manufacturers had withdrawn at least one device from the EU market since 2021. The most frequently cited reason: revenue does not cover the cost of re-approval. That is not a safer patient. That is a patient the device never reaches.

A disclaimer, to be fair

Technical regulation has its paperwork as well. EN 60601-1 requires risk management and, for programmable systems, refers to a process standard for software development. Hardware in the higher classes also goes through the same MDR as software. Conversely, software can be measured: the MDCG 2020-1 guidance explicitly asks for technical and clinical performance to be demonstrated on data. The difference lies in what actually decides conformity in practice. For current, it is the meter. For software, it is the binder.

A telling detail: since June 2026, EN 60601-1 has been harmonised under the MDR, so meeting it gives a presumption of conformity. The key software life cycle standards are not on that list yet.

I am not saying administration is pointless, and certainly not that regulation is. Only that administration has a price, and the debate tends to leave it out.

Who can carry it

A large manufacturer absorbs it in the budget. It has a quality department, regulatory specialists and a portfolio to spread the costs across. A small team, or a person who knows the problem from the inside, often stops before they start. And yet it is precisely such people who come up with solutions a large manufacturer cannot see from a distance.

If even part of that energy shifted from paper to measurement, meaning test sets, performance verification and work with clinicians, I do not think safety would suffer. If anything, the opposite.

So where is the line beyond which administration no longer adds safety, only cost and delay?